MIC-10k1 & MIC-5050 Insulation Resistance Meters

Key Features:
- digital filters function for measurements in high noise environment,
- continuity measurement of protective connections and equipotential
bonding in accordance with EN 61557-4 with current > 200 mA,
- adjustable limits for measured resistance R and R , ISO CONT
- measurement of leakage current during insulation resistance testing,
- DC and AC voltage measurement in the range of 0...750 V,
- drawing graphs on the display during measurement,
- innovative memory with possibility of description of: measurement points,
facilities, names of customers.
- operating with mini Bluetooth keyboard (option),
- graphic LCD 5,6'’ backlit,
- keyboard backlit,
- power supply from main power line or battery packs,
- built-in fast charger,
- the instruments meet the requirements of the EN 61557 standard.
Insulation resistance measurement:
· up to 40 TΩ for MIC-10k,
· up to 20 TΩ for MIC-5050,
- measurement voltage any in the range of:
· 50...5000 V for MIC-5050 (50...1000 V at 10 V and
1...5 kV at 25 V),
· 50…10000 V for MIC-10k1 (50...1000 V at 10 V and
1...10 kV at 25 V),
- continuous indication of measured insulation resistance or leakage
current,
- automatic discharge of measured object capacitance voltage after
the end of insulation resistance measurement,
- acoustic signaling of 5 seconds intervals to facilitate capturing time
characteristics,
- adjustable measuring time to 99'59'’,
- metered T , T and T test times for measuring one or two absorption
coefficients from the range of 1…600 s, 1 2 3
- polarization index (PI), absorption coefficients Ab1, Ab2 and
dielectric absorption ratio (DAR) measurement,
- indication of actual test voltage during measurement,
- 1.2 mA, 3 mA and 5 mA test current,
- insulation resistance measurement using two- or three-wire
method,
- measurements with test leads up to 20 m...
- protection against measuring live objects,
- automatic measurement of multiple core cables with the additional
AutoISO-5000 adapter (for MIC-10k1 max. voltage 5 kV)
- measurement of capacitance during the measurement of R . ISO
- measurement of temperature (with additional probe - WASONT1),
- step voltage insulation resistance measurement (SV),
- Dielectric Discharge calculation (DD),
- location of damage (burnout),
Specifications:
Electrical safety
- type of insulation double, EN 61010-1 and IEC 61557 compliant
- measurement category IV 600 V (III 1000 V) according to EN 61010-1
- degree of housing protection acc. to EN 60529 IP40 (IP67 for closed enclosure)
Other technical specifications
- power supply of the meter ....................................................................12 V gel battery
90 V ÷ 260 V 50 Hz/60 Hz from the electric grid
- dimensions .................................................................... 390 mm x 310 mm x 180 mm
- meter weight approx............................................................................................... 7 kg
- storage temperature ............................................................................... -25˚C...+70˚C
- working temperature .............................................................................. -20˚C...+50˚C
- humidity ...................................................................................................... 20%...80%
- altitude (above sea level)............................................................................... <3000 m
- reference temperature .......................................................................... +23 ˚C ± 2 ˚C)
- reference humidity........................................................................................ 40%...60%
- display ..................................................................................... LCD 5,6', segment-type
transmission of measurement results ................................................. USB or Bluetooth
- number of R measurements, acc. to EN 61557-2 with battery power supply min. 1000
- quality standard........................ design, construction and manufacturing are ISO 9001,
ISO 14001, PN-N-18001 compliant
- the device meets the requirements of the EN 61010-1 and IEC 61557 standards
- the product meets EMC requirements (immunity for industrial environment) according to
.*For full Technical Specifications, please refer to PDF Data Sheet
Standard Accessories:
-
"crocodile" clip 5,5 kV; black
-
"crocodile" clip 5,5 kV; red
-
"crocodile" clip 5,5 kV; blue
-
pin probe 5,5 kV with banana connector; red
-
pin probe 5,5 kV with banana connector; black
-
power cord
-
temperature probe ST-1
-
battery pack (built-in)
-
test lead banana plug; 3 m; 10kV; red
-
test lead „E” banana plug; 3 m; 10 kV; blue
-
test lead banana plug; 1,8 m; 10 kV; black; shielded
-
USB cable
-
carrying case L4 for accessories
-
“SONEL Reader” software
-
calibration certificate