ΩMEGASAFE EP 4Ex Ex-proof low-impendance tester
Features & Functions
Resistance Measuring:
-
Precise indication of resistance between 10 mΩ up to 2 MΩ
-
Universally applicable due to large measuring range and automatic range selection
-
Energy-saving circuit for high measuring current 0.2 A for low-ohm range up to 10 Ω
-
Large display with backlight
-
Acoustic warning signal for resistances less than 2 Ω
-
Fast compensation of measuring line per keypress
-
Automatic pole change and storage of measurement results (+/- values)
Safety:
-
Clear LED indicator for dangerous voltage
-
Voltage value displayed on LCD with warning acoustic signal
-
Protection circuit against external voltages up to 400 V
-
Integrated fuse for overvoltages > 400 V
Further Functions:
-
Varied accessories
-
Easy handling and compact design
-
Robust casing
-
IP 65 – waterproof
Meet The ΩMegaSafe EP 4Ex
-
resistance range: 0,01 Ohm – 2 MOhm
-
safety: CAT III 300 V / CAT II 600 V
-
Ex Protection II 2G Ex ib IIC T4
-
including test line

Optional Accessories:

Technical data
Model:
EP 4
EP 4Ex
ISO 1
ISO 1Ex
Resistance Range
-
EP 4: 0.01 Ohm – 2 MOhm
-
EP 4Ex: 1 kOhm – 20 GOhm
-
ISO 1: 1 kOhm – 2 TOhm
-
ISO 1Ex: 1 kOhm – 2 TOhm
Test Voltage
-
EP 4: -
-
EP 4Ex: 100 V, 250 V, 500 V, 1000 V
-
ISO 1: 32V, 100 V, 500 V
-
ISO 1Ex: 32V, 100 V, 500 V
Test Current
-
EP 4: max. 200 mA
-
EP 4Ex: 1 mA
-
ISO 1: -
-
ISO 1Ex: -
Continuity Tester with Sound
-
EP 4: Yes
-
EP 4Ex: No
-
ISO 1: No
-
ISO 1Ex: No
Connection / Electrode
-
EP 4: + Electrode 4 mm, - Socket for 4 mm plug
-
EP 4Ex: + Electrode 4 mm, - Socket for 4 mm plug
-
ISO 1: + Electrode 4/15 mm, - Socket for 4 mm plug
-
ISO 1Ex: + Socket for 4 mm plug, - Socket for 4 mm plug
Power Supply
-
All models: 1 x 9 V
Protection Class
-
All models: IP 65
Ex Approval According to ATEX
-
EP 4: No
-
EP 4Ex: Yes, ll 2G Ex ib llC T4 Gb (according to DIN EN 60079-0 and DIN EN 60079-11)
-
ISO 1: No
-
ISO 1Ex: Yes, ll 2G and l M2 ib llB T4 (according to DIN EN 60079-0 and DIN EN 60079-11)
Overvoltage Category
-
EP 4, EP 4Ex: CAT III 300 V, CAT II 600 V (according to EN/IEC 61010-1)
-
ISO 1, ISO 1Ex: CAT IV 600 V, CAT III 1000 V, CAT III 600 V (according to EN/IEC 61010-1)
Ex Maximum Values
-
EP 4, ISO 1: -
-
EP 4Ex: U0 = 8.7V, I0 = 220 mA, P0 = 1.3 W
-
Maximum allowable inductance: 0.2 mH
-
Maximum allowable capacitance: 5.0 µF
-
-
-
ISO 1Ex: U0 = 550 V, I0 = 1.5 mA
-
Maximum allowable inductance: 28 H
-
Maximum allowable capacitance: 1 nF
-
-
Ambient Temperature
-
EP 4, ISO 1: -10 ... +50 °C
-
EP 4Ex, ISO 1Ex: -10 ... +40 °C
Application Area
-
EP 4, EP 4Ex: Testing protective measures according to IEC 61557-4 (VDE 0413 Part 4), e.g., protective conductor testing or machine grounding testing
-
ISO 1, ISO 1Ex: Testing protective measures according to IEC 61557-2 (VDE 0413 Part 2), e.g., leakage testing of floors and surfaces
Article Number
-
EP 4: 83302
-
EP 4Ex: 83301
-
ISO 1: 83315
-
ISO 1Ex: 83310
Product Dimensions & Weight
-
EP 4: 230 x 60 x 40 mm, 180 g
-
EP 4Ex: 230 x 60 x 40 mm, 220 g
-
ISO 1: 230 x 60 x 40 mm, 220 g
-
ISO 1Ex: 230 x 60 x 40 mm, 220 g
Packaged Product Dimensions & Weight
-
EP 4: 300 x 100 x 50 mm, 350 g
-
EP 4Ex: 300 x 100 x 50 mm, 370 g
-
ISO 1: 300 x 100 x 50 mm, 400 g
-
ISO 1Ex: 300 x 100 x 50 mm, 450 g