Introduction to the  MD5060x:
  • The TENTECH MD5060x is a smart, microprocessor-controlled, 5 kV insulation tester. Besides the conventional measurement of insulation resistances up to 5 TΩ, its features allow to automatically measure both the polarization index and dielectric absorption index, thus significantly simplifying testing of transformers.
     

  • This insulation tester is a portable, battery-powered equipment. Test voltage may be chosen from 500 V up to 5 kV in 100 V increments. Due to its measurement principle (actual voltage and current readings) the accuracy of resistance measurement is not affected by any test voltage error. Nevertheless, the high-voltage generator is electronically stabilized with a better than 2 % accuracy in open-circuit condition and for any normal operation battery voltage.
     

  • The built-in chronometer automatically counts the elapsed time since the start of measurement. Measured values are transmitted through the USB interface and are printed in the built-in printer as a registration of the performed test. Furthermore, the measured values are stored in a non-volatile internal memory. Up to 4000 measured values may be stored to be transferred afterward to a computer running the T-Logger program. This software allows a further analysis of the test results, including a graphical representation and automatic report generation. The real time clock and calendar, and the sequential test number facilitates the identification of each test and the organization of a predictive maintenance system by trend analysis.
     

  • Some other advanced features are useful to run the most sophisticated insulation analysis. Step voltage test, configurable pass / fail test and timed measurements are automatically performed, with a very simple and user-friendly setup. The measurement parameters are stored in the non-volatile memory for an easy configuration.
     

  • In order to guarantee the operator safety, the cabinet is made of a high breakdown strength plastic. High security output terminals are placed in the furthest and most protected area of the equipment. A light indicator shows high-voltage presence and is switched-off when the capacities (both internal and external) are discharged up to reaching a safe value.
     

  • The MD5060x is powered by a rechargeable LFP battery and has an automatic charger that may be connected to 100 - 240 V mains supply (AC Adapter). The cabinet is strong and lightweight, easy to carry, impact-resistant and suitable to be used under severe weather conditions. Thus the megohmmeter supplies very reliable and accurate measurements both in laboratory and out in the field.

The  MD5060x
Technical Specifications

Test voltages

  • 500, 1000, 2500, 5000 V directly, one button selectable

  • 500 V to 5 kV in 100 V increments

  • DC, negative

Maximum resistance reading

  • 5 TΩ @ 5 kV

Short circuit current

  • 1.5 ± 0.5 mA

Test voltages accuracy

  • ± 3 % of nominal test voltages on R ≥ 10 GΩ

Insulation tester basic accuracy

  • ± 5 % of reading from 1 MΩ up to 1 TΩ

  • ± 20 % of reading from 1 TΩ up to 5 TΩ

Advanced features

  • Automated polarization index calculation

  • Automated dielectric absorption ratio calculation

  • Step voltage test

  • Programmable pass/fail test

Built-in printer

  • Prints elapsed time, actual voltage, and resistance measured every 15 seconds

Built-in memory

  • Stores up to 4000 measured values

Interface

  • USB

Built-in chronometer

  • Shows elapsed time in mm:ss format

  • Count starts automatically for each measurement

Real-time clock and calendar

  • Allows a reliable identification for each test

T-Logger software

  • Friendly, easy-to-use software

  • Tests are represented in graphical and tabular views

  • Includes automatic report generator with operator comments

Environmental protection

  • IP54 (with closed lid)

Safety class

  • In accordance with IEC 61010-1

E.M.C.

  • In accordance with IEC 61326-1

Electrostatic immunity

  • In accordance with IEC 61000-4-2

Electromagnetic irradiation immunity

  • In accordance with IEC 61000-4-3

Power supply

  • Internal rechargeable LFP 12 V - 3000 mAh battery

Battery charger

  • 12 V - 2 A

Operating temperature range

  • 14 °F to 122 °F (-10 °C to 50 °C)

Storage temperature range

  • -13 °F to 158 °F (-25 °C to 70 °C)

Humidity range

  • 95 % RH (non-condensing)

Equipment weight

  • Approx. 7.94 lb (3.6 kg)

Dimensions

  • 10.78" x 9.84" x 4.88" (274 x 250 x 124 mm)

Features
  • Insulation resistances up to 5 TΩ

  • Auto-range

  • Guard terminal

  • Digital & bargraph reading

  • Alphanumerical display

  • Programmable timer

  • Automated tests: absorption index and polarization index

  • Step voltage test

  • Programmable pass-fail

  • Real time clock and calendar

  • Built-in printer

  • 4000 readings memory

  • USB interface

  • Rechargeable LFP battery

  • Auto power-off

  • Software for data management

Included accessories
  • 2 Measuring test leads.

  • 1 GUARD test lead.

  • 1 AC Adapter.

  • 1 T-Logger software (CD-ROM).

  • 1 User guide.

  • 1 Protective bag.

  • 1 USB cable.

Desktop software
  • Equipment's internal memory can be accessed by a desktop software that allows to download all memory data. It also can be used to make analysis of the test results and automatic report generation. (Software included with purchase)

pic_notebook.png